By optimizing the bridge circuit design and temperature control algorithm, the AC Bridge Temperature Controller overcomes the challenges of high signal noise and slow control response in traditional cryogenic measurements. It provides excellent temperature measurement and control capabilities for dilution refrigerators (DRs) operating below 100 mK. It is particularly suitable for quantum device characterization scenarios that require long-term, stable maintenance of low temperatures in the mK range.
Key Features:
Parameters | Specifications |
---|---|
Feature | Supports dilution refrigerator (DR) operation below 100 mK |
Up to 16 measurement channels using an optional scanner | |
Powerful impedance measurement capabilities | |
Multiple PID-controlled outputs, heater power up to 10 W | |
Input type | AC, 4-line difference, resistance |
Channel | 1,Maximum16 Channel (a 16-channel scanner is required,Optional) |
Resistance range | 22 Range, from 2mΩ to 63.2 MΩ (depending on the incentive) |
Temperature coefficient | ± 0.0015% / (Relative Read) |
Excitation power supply | Sinusoidal AC current source |
Excitation current | 22 Range, ranging from 1 pA to 31.6 mA (valid value) |
Excitation accuracy | ±2% |
Minimum excitation power | 10E-18W at 100 kΩ |
Voltage input range | 12 Range, from 2 μV to 632 mV (valid value) |
Range selection mode | Manual, voltage excitation, current excitation, automatic range |
Scanner mode | Manual or automatic scanning(Optional) |
Filter | Stabilization time : 1s~200s filter window : 1%~80% |
Supported sensors | NTC/PTC resistance sensor |
Connection | 6 Needle DIN (current output), 6-pin DIN (voltage input), DA-15 (Scanner control) |
Address: No. 1, Jikang Road, Kengzi Subdistrict, Pingshan District, Shenzhen, Guangdong Province
Hotline: +86 0755-28329695 / +86 0510-85590779