Products

KECV-325 Electrochemical C-V Profiler

KECV‑325 is an electrochemical capacitance‑voltage profiler. It etches solar cell wafers via combined circuitry and chemical electrolyte. Through cooperation of upper computer and lower controller, it calculates carrier concentration of each layer of the cell wafer, deduces carrier distribution after diffusion process, and clearly evaluates diffusion process quality. It serves as key characterization equipment for photovoltaic and semiconductor materials & devices.

Technical Specifications
Item Parameters
Measurement Type Doping profile (active carrier concentration), capacitance‑voltage‑based depth profiling
Applicable Materials Doped semiconductors (n‑type / p‑type), especially suitable for phosphorus‑doped and boron‑doped layers
Depth Resolution Near‑surface (nm) to several μm; depth range from 10 nm to 10 μm (depends on material and doping)
Concentration Range Active carrier concentration: 1×10¹⁴ ~ 5×10¹⁷ atoms/cm³ (wide dynamic range)
Layer Types Polycrystalline silicon (n+, p+ or p), crystalline silicon, tunnel oxide interface: ideal choice for post‑TOPCon & selective emitter profile characterization
Etch Rate Control Real‑time depth control (etch rate normally calibrated against reference materials)
Data Output Carrier concentration‑depth plot output in logarithmic or linear scale
Measurement Speed 10–50 minutes per point, subject to layer thickness and doping
Automation Workstage movement, electrolyte dispensing, semi‑automatic / fully automatic wafer mapping & data logging
Sample Compatibility Wafer size from 182 mm² to 230 mm²; custom PV wafer chuck available
Safety Features Acid container, leakage control, emergency stop including waste treatment & alarm
Software Functions Real‑time data visualization, CSV / Excel export, overlay & batch comparison of multiple calibration datasets, statistics
Common Electrolytes HF, HCl, CH₃COOH; custom etchant formulation based on dopant & material type